STEP 4 : Check if the Test Result is in Range.
After seven (7) seconds, the control solution test
result appears on the display. Compare the test
result with the range printed on the test strip vial.
The result should fall within the printed range.
See Figure 21.
Out of Range Results
If test results fall outside the range
printed on the test strip vial, check
“Trouble-Shooting Guide” located in the
“System Troubleshooting” Section and
repeat the test.
If you continue to get out-of-range results, it means that the system
or the control solution may not be working properly. DO NOT use the
system to test your blood glucose level. If you are unable to resolve
the problem, contact Technical Support at 1-800-243-2636.
NORM AL
109-165
Figure 21