Prodigy Diabetes Care, LLC www.prodigymeter.com
PRODIGY
®
Blood Glucose Meters Owner’s Manual
70
71
Prodigy Diabetes Care, LLC 1.800.243.2636
PRODIGY
®
Blood Glucose Meters Owner’s Manual
Cleaning your System
Your Prodigy
®
meter should be cleaned whenever it is visibly dirty by
wiping the outside of the meter using a cloth dampened with either
mild detergent mixed with water or 70% isopropyl (rubbing) alcohol.
Do not use bleach or other harsh abrasives to clean your meter.
Do not get any liquids, dirt, dust, blood or control solution inside the
meter through the test strip port or the data port. Never spray cleaning
solution on the meter or immerse it in any liquid.
Lancing Device
Wipe your Prodigy
®
Lancing Device with a soft cloth dampened with
water and mild detergent or 70% isopropyl (rubbing) alcohol. Do not
immerse the lancing device in any liquid. Do not use bleach.
Disinfecting your System
If your Prodigy
®
meter is being operated by a second person who is
providing testing assistance to the user, the meter and lancing device
should be decontaminated prior to use by the second person.
Lancing Device
Disnifect your Prodigy
®
Lancing Device about once a week. After
cleaning the device, unscrew the cap and place it in 70% isopropyl
(rubbing) alcohol for several minutes. Make sure the cap is completely
dry before reattaching it to the device.
Please contact Customer Care if you have any questions at
1.800.243.2636.
Caring for your Prodigy
®
Meter
Storing Your System
The Prodigy
®
meters do not require special maintenance.
• Store the meter, test strips and control solution in the carrying
case after each use in a cool, dry place.
• Do not refrigerate.
• Do not store in the kitchen or bathroom.
*Your meter is a precision instrument. Please handle it with care
.
Battery
The Prodigy Voice
®
and Prodigy AutoCode
®
meters come with two (2)
1.5 V AAA alkaline batteries. The Prodigy Pocket
meter comes with
one (1) CR2032 battery. Batteries are manufactured in a charged state
and not designed for recharging. Recharging the batteries can cause
battery leakage, or in some cases, high pressure rupture. Replace all
batteries at the same time.
Warning: Batteries can explode or leak and cause burns if
installed backwards, disassembled, charged or exposed to water,
reorhightemperatures.
Low Battery
The meter will alert you when the power is getting low by displaying two
(2) different messages:
1. When the “ ” symbol appears alone on the display, the meter
is functional and the result remains accurate, but you should
change the battery as soon as possible.
2. When the “ ” symbol appears together with the “E-b” symbol
on the display, the battery does not have enough power for a test.
You must change the battery before using the meter.